Register | Login
  A21 Gas Sampler
 
Operational and Maintenance Manual(s)
are available to Registered Users.
 
Registered Users login HERE.
 
To Register click HERE.

The model A21 is a gas sample pumping system designed to draw samples from ducts or hard to access locations and deliver the samples to gas sensors located in a more convenient area. A21 Sampling Systems use a high quality diaphragm pump operated by a brushless DC motor to provide continuous sampling over long periods of time.

Sampling pumps are rated for over 10,000 hours of continuous operation and diaphragms in the pump may be changed easily if necessary. The pump will draw samples against a vacuum of up to 10”Hg. for sampling negative pressure duct systems when necessary.

In addition to the sampling pump, A21 systems contain various components needed to insure reliable sample measurement. A flowmeter mounted on the front panel allows the user to adjust the sample flowrate to the desired value, normally around 500 cc./min. Internal to the system is a loss of flow detector that will warn of sampling system problems that could compromise the gas measurement system. A loss of flow alarm light on the front panel provides local flow loss indication while an internal relay can be used to indicate this alarm remotely.

  Gas Sample Pumping System (Cut / Paste)

Click the link below
to download the typical specs
in Microsoft® Word Format:

   A21 Gas Sampling System

A Gas Sample Pumping system shall be supplied for the purpose of extracting gas samples from air ducts, vent stacks, enclosed spaces or other areas where direct gas detection and measurement might be difficult. The Gas Sampling System shall be ATI Series A21 as described below.

The Gas Sample Pumping system shall be designed to continuously draw samples through tubing at a flow rate up to 1000 cc/min and to be adjustable at rates between 300 – 1000 cc/min via a rotameter located on the front panel. The Gas Sampling system shall pump against a maximum 10” Hg. on the inlet or 4 PSIG pressure on the outlet for 500 cc/min. and provide a response through 1/8” tubing at 29 seconds/ 100 feet, 3/16” tubing at 65 seconds/ 100 feet and 1/4” tubing at 116 seconds / 100 feet. A low flow alarm lamp shall be visible from the front panel and include a SPDT relay to indicate loss of sample flow.

The Gas Sample Pumping system shall be powered by 120 VAC or 12 VDC (specify one power source only).



 
 
Get Adobe® Reader® Get Microsoft® Word Viewer
   

Click above link to download Adobe Reader.

Click the above link to download Microsoft Word Viewer for Windows® if you do not already have Microsoft Word.
Copyright (c) 2001 - 2010 by ATI